ALINITY C PROCESSING MODULE AND ACCESSORIES

Category:
FEATURE ALINITY ALINITY i ALINITY ci
Dimension (H x W x D) 135 x 119 x 117 cm/1.39 135 x 119 x 117 cm/1.39 135 x 200 x 117 cm/2.33
Methods Photometric, Potentiometric Chemiluminescence Photometric, Potentiometric, Chemiluminescence
Thoughput Up to 1350 TPH Up to 200 TPH Up to 1550 TPH
Scalability Up to 971 TPH/ Up to 144 TPH/ Up to 664 TPH/
Continuous Access of Reagents, Calibrators, Controls, and Consumables Yes Yes Yes
Flexible Stat Options Prioritize single rack as needed or configure multiple fixed positions
Sample Types Serum, plasma, urine, cerebrospinal fluid, hemolysate, whole blood Serum, plasma, whole blood, urine Serum, plasma, urine, cerebrospinal fluid, hemolysate, whole blood
Sample Capacity 150 150 300
Sample Bar Code Types Code 128, standard code 39, interleaved 2 of 5, codabar Code 128, standard code 39, interleaved 2 of 5, codabar Code 128, standard code 39, interleaved 2 of 5, codabar
Sample Result Storage 200,000 200,000 200,000
Dead Volume 50 µL (sample cup) 50 µL (sample cup) 50 µL (sample cup)
Sample Volume+ 1.5-35µL 2-200µL For Anality c: 1.5-35µL

For Anality i:  2-200µL

Sample Probe Carryover ≤0.1 parts per million** ≤0.1 parts per million** ≤0.1 parts per million**
Reagent capacity Up to 70 refrigerated reagent cartridges on-board plus patented ISE (NA+, K+, and CI-) Up to 47 refrigerated reagent cartridges onboard Up to 117 refrigerated reagent cartridges onboard plus patented ISE (NA+, K+, and CI-)
Reagent Type 100% liquid ready-to-use 100% liquid ready-to-use 100% liquid ready-to-use
Reagent On Board Stability+ 5-65 days 15-30days For Anality c: 5-65days

For Anality i: 15-30 days

Automated Onboard Calibrators and Controls+ Yes Yes (controls only) Yes
Calibration Frequency+ 1-60 days 15-30 days For Anality c: 1-60 days

For Anality i: 15-30 days

Sample, Clot and Bubble Detection Yes Yes Yes
Reagent Pressure Monitoring In Development Yes For Anality c: In Development

For Anality i: Yes

Sample Interference Measurement Yes-hemolysis, icterus, and lipemia No Yes-hemolysis, icterus, and lipemia (CC Only)
On Board Maintenance Records Yes Yes Yes
Online Error Code Help Yes Yes Yes
Host Interface HL7 or ASTM HL7 or ASTM HL7 or ASTM
Remote Diagnostics AbbottLink AbbottLink AbbottLink
Weight 712 kg 624kg 1160kg
Electrical Requirements AC 180-264 V, 20 Amp AC 180-264 V, 30 Amp AC 208-240 V, 20 Amp + AC 200-264 V, 30 Amp
Water Requirements Average <27 L/hr: <30 L/hr max flow rate Average <10 L/hr: <30 L/hr max flow rate Average <27 L/hr: <30 L/hr max flow rate (CC)

Average <10 L/hr: <30 L/hr max flow rate (IA)

Heat Output Average 3481Btu 1634 Btu (typical running mode) 5115 Btu
Noise Level-operator standing in front system (0.25m) Does not exceed 55.7 dB Does not exceed 65.1 dB For Anality c: Does not exceed 55.7 dB

For Anality i: Does not exceed 65.1 dB

Laboratory Automation Connection In Development In Development In Development

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.

Discussions

There are no discussions yet.

Leave a reply

Your email address will not be published. Required fields are marked *